9.7 Message Reference
135
9
Delimiter Setting
Syntax
Command
:SYSTem:TERMinator
<0/ 1>
Query
Response
:SYSTem:TERMinator?
<0/ 1>
<0>
= LF+EOI
<1>
= CR+LF with EOI
Example
:SYST:TERM 1
:SYST:TERM?
0
Note
•
At power-on, this is set to 0 (LF+EOI).
•
The RS-232C terminator is fixed as CR+LF.
Set and Query the System Date
Syntax
Command
:SYSTem:DATE
<Year>
,
<Month>
,
<Day>
Query
Response
:SYSTem:DATE?
<Year>,<Month>,<Day>
<Year>
= 00 to 99
<Month>
= 01 to 12
<Day>
= 01 to 31
Description
Sets and queries the date of the real-time system clock.
Error
Attempting to set an out-of-range numerical value returns an execution error.
Attempting to set a non-existent date (such as 09,06,31) returns an execution error.
Example
:SYST:DATE 9,10,5
:SYST:DATA?
09,12,03
Set and Query the System Time
Syntax
Command
:SYSTem:TIME
<Hours>
,
<Minutes>
,
<Seconds>
Query
Response
:SYSTem:TIME?
<Hours>,<Minutes>,<Seconds>
<Hours>
= 00 to 23
<Minutes>
= 00 to 59
<Seconds>
= 00 to 59
Description
Sets and queries the time of the real-time system clock.
Example
:SYST:TIME 08,25,00
:SYST:TIME?
23,09,53
Measurement Settings Backup
Syntax
Command
:SYSTem:BACKup
Description
Stores current measurement settings in non-volatile memory. Changes to measurement settings
by communications commands are not backed up (and are lost when power is turned off).
Use this command to store the settings as occasion demands.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......