8.6 Supplied Connector Assembly
96
The EXT I/O connector and shell are supplied with the instrument. Assemble as shown below.
• Use shielded cables to connect a PLC to the EXT I/O connector. Using non-shielded conductors may result
in system errors from electrical noise.
• Connect the shield to the ISO_COM pin of the EXT I/O connector.
8.6
Supplied Connector Assembly
E
Nut #4-40UNC
Required tools:
•
Screwdriver
•
Shielded cable
•
Soldering iron
Accessories
•
A
Cover.....................................2
•
B
Screws (+/-) #4-40UNC
(16.9 mm overall length) .......2
•
C
Screws (+/-) #4-40UNC
(12.6 mm overall length) .......2
•
D
Screws (
) #4-40UNC
(15.0 mm overall length) .......2
•
E
Nuts #4-40UNC .....................2
•
F
Cable Clamps........................2
•
G
Saddle Washers
(shell protection)....................2
•
H
Connector ..............................1
B
Screw (+/-) #4-40UNC (16.9 mm overall length)
A
Cover
(Identical halves)
Shielded Cable
(recommended)
D
Screw (
) #4-40UNC
(15.0 mm overall length)
B
Screw (+/
) #4-40UNC
(16.9 mm
overall length)
E
Nut #4-40UNC
D
Screw (
) #4-40UNC
(15.0 mm overall length)
A
Cover
(Identical halves)
F
Cable Clamps
G
Saddle Washers
(shell protection)
Assembly Sequence
1. Solder the (shielded) cable wires to the supplied EXT I/O connector (H) pins.
2. Affix the cable clamps (F) on the cable with screws (C).
3. Position the cable clamps (F) to fit properly inside the cover (A).
4. Insert screws (D) through the saddle washers (G).
5. In one half of cover (A), place connector (H), clamps (F), saddle washers (G) and screws (D).
6. Place the other half of cover (A) on top.
7. Affix the halves of the cover (A) together with screws (B) and nuts (E).
Be careful not to overtighten the screws, which could damage the covers.
H
Connector
C
Screw (+/
) #4-40UNC
(12.6 mm overall length)
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......