Appendix 9 Zero Adjustment
A
18
To perform zero adjustment using a probe
When you actually perform zero adjustment using a probe, you may unexpectedly
make the connection shown in Table 1 (b). Therefore, when performing zero
adjustment, you need to pay sufficient attention to the connection state of each
terminal.
Here, 9287-10 Clip Type Lead is used as an example for the connection explanation.
Table 2 shows the connection state of the tip of the lead and equivalent circuit in the
respective correct and wrong connections. Table 1 (a) indicates the correct
connection method, resulting in 0 V between H
POT
and L
POT
. However, Table 1 (b) is
the wrong connection method, so that 0 V is not obtained between H
POT
and L
POT
.
Table 2: Clip type lead connection methods used during zero adjustment
Connection method
Correct
Wrong
Tip of lead
Equivalent circuit
Deformed equivalent
circuit
As connection
method for zero
adjustment
Correct
Wrong
Red
Black
Red
Black
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......