9.4 Configuring the Communications Protocol
101
9
Configure the interface on the System screen.
External command operation is undetermined when the Printer is selected as the interface type. In
this case, commands should not be sent.
Make these instrument settings.
Configure the controller (PC or PLC).
9.4
Configuring the Communications Protocol
Configuring RS-232C Interface Communications
1
Open the Basic Settings screen.
2
Open the System screen.
3
Select the interface type.
4
Select the interface transfer rate.
5
Return to the Measurement screen.
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
2
RS-232C Interface
1
Selection
2
9600 (bps)(default)
19200 (bps)
38400 (bps)
1
The confirmation screen appears.
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
Be sure to make set up the controller as shown below.
•
Asynchronous communication
•
Transfer rate: 9600bps/ 19200bps/ 38400bps
(set to match the instrument setting)
•
Stop bit: 1
•
Data length: 8
•
Parity check: None
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......