9.7 Message Reference
125
9
Read/Write the Standard Event Status Enable Register (SESER)
Syntax
Command
ESE
<0 to 255 (NR1)>
Query
Response
ESE?
<0 to 255
(NR1)>
Description
Command
The SESER mask is set to the numerical value 0 to 255.
The initial value (at power-on) is 0. ("Default Settings" (p. 68))
Query
The contents of the SESER, as set by the
ESE
command, are returned as an NR1
value (0 to 255).
Example
ESE 36
(Sets bits 5 and 2 of SESER)
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Read and Clear the Standard Event Status Register (SESR)
Syntax
Query
Response
ESR?
<0 to 255
(NR1)>
Description
Returns the contents of the SESR as an NR1 value from 0 to 255, then clears register contents.
The response message has no header.
Example
ESR?
32
Bit 5 of the SESR was set to 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
unused
CME
EXE
DDE
QYE
unused
unused
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......