5.7 Initializing (Reset)
68
Default Settings
Display
Setting value
Default Settings
Setting Description
MENU
TRG
INT/ EXT/ MANU
EXT
Trigger source selection (p. 30)
RANGE
/
[Low Power: Off] 100m
/
1000m
/ 10
/ 100
/ 1000
/
10k
/ 100k
/ 1000k
/ 10M
/
100M
[Low Power: On] 1000m
/ 10
/
100
/ 1000
100 M
Range selection (p. 31)
SPEED
SLOW/ MED/ FAST
FAST
Measurement speed (p. 29)
0ADJ
OFF/ ON
OFF
COMP
OFF/ ON
ON
Comparator function(p. 34)
LOCK
OFF/ FULL / MENU
OFF
MISC
MEAS/ DATA/ SYSTEM
(Miscellaneous settings)
MISC
MEAS
DELAY1
0 to 100 ms (all ranges)
0 ms
DELAY2
0 to 100 ms
0 ms
DUT response setting (p. 42)
INT (FAST)
0.1 ms to 100 ms
1PLC to 6PLC (60 Hz)
1PLC to 5PLC (50 Hz)
Depends on measurement range
Integration time (p. 44)
INT (MED)
INT (SLOW)
CONT CHECK
OFF/ ON
ON, 200
Contact-check (p. 46)
50
/ 100
/ 150
/ 200
/ 300
/ 400
/ 500
CONT IMP
OFF/ ON/ PULSE
17 mA/ 25 mA/ 35 mA/ 50 mA
ON, 35 mA (range from 100 m
to
100 k
)
PULSE, 35 mA (range from1000
k
to 100 M
)
Contact Improvement (p. 47)
VOLT
MONITOR
OFF/ ON
ON, NORMAL (LOOSE when
100
Range)
Voltage level monitor (p. 49)
LOOSE/ NORMAL/ SEVERE
CURRENT
MODE
CONT/ PULSE
PULSE
Current mode setting (p. 50)
DATA
AUTO MEMO-
RY
OFF/ ON
OFF
Auto-Memory function (p. 71)
STATISTICS
OFF/ ON
OFF
Statistical calculation function
(p. 74)
DATA OUT
OFF/ ON
OFF
SYS-
TEM
SET MONITOR
OFF/ ON,
1st/ 2nd, 0.000% to 9.999%
OFF, 1st, 0.000%
Settings Monitor function (p. 53)
PROBE
CHECK
OFF/ ON, 0 to 100 ms
OFF, 5 ms
Probe short-circuit detection (p. 51)
RETRY
OFF/ ON, 1 to 50 ms
ON, 50 ms
TRIG EDGE
OFF EDGE/ ON EDGE
ON EDGE
EOM
PULSE/ HOLD
1 to 100 ms
PULSE, 5 ms
End-of-measurement pulse width
(p. 93)
INTERFACE
GP-IB/ RS232C/ PRINT
RS232C, 9600bps
GP-IB, ADR01, LF
Interface setting (p. 101)
LOW POWER
OFF/ ON
OFF
Low-Power Resistance Measure-
ment (p. 28)
JUDGE BEEP
OFF/ ON
IN/ HI/LO/ LOW/ HIGH
OFF, HI/LO
Comparator decision beeper (p. 61)
KEY BEEP
OFF/ ON
ON
Key beeper (p. 59)
CLOCK
LINE FREQ
AUTO/ 50 Hz/ 60 Hz
AUTO
Line frequency (detection) (p. 63)
CONTRAST
0 to 100
50
Screen contrast adjustment (p. 65)
BACK LIGHT
0 to 100
80
Screen backlight adjustment (p. 66)
RESET
-
-
ADJUST
-
-
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......