4.2 Setting Pre-Measurement Delay
42
This setting specifies the delay between trigger signal input and the start of measurement.
Adjust this setting to delay measurement until the measured value has time to stabilize, so that even
if the sample is connected after triggering, measurement starts only after the specified delay. The
delay can be set by two methods, as follows.
4.2
Setting Pre-Measurement Delay
The DELAY1 setting is common to all ranges. The default set-
ting is 0 ms (corresponding to trigger signal input at the same
time as probe contacts become stable). Setting DELAY1 affects
measurements in all ranges.
Set DELAY1
Adjust this setting to allow for
probe contact mechanical stabili-
zation.
Set DELAY2 to the time needed for stabilization after measure-
ment current is applied, such as may be required for inductive
components. The setting affects only the selected range. The
default setting is 0 ms (corresponding to resistance measure-
ment of non-inductive components).
Set DELAY2
Adjust this setting to allow for sta-
bilization of the measurement
sample.
DELAY1 and DELAY2 Timing Chart
* Internal delay is provided to suit purely resistive (non-reactive) DUTs,
and is different for each measurement range.
Acquisition
Stable Contact
Internal
Delay*
Probe Contact
Condition
Start
TRIG
Measurement
Current
Acquisition of
Measured Value
End of Measurement
Signal
EOM
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......