4.8 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
51
4
Four-terminal measurements are not possible when a conductive for-
eign object is present between the POT and CUR probe tips. To
detect short-circuited probes, this function measures the resistance
between the CUR and POT terminals after a specific time (initially 5
ms) following the end of measurement. Probe short-circuit detection
is disabled by default.
When a probe short-circuit is detected, an error message appears on
the measurement screen, and the
PRB_SHORT
and
ERR
signals are
output. (
ERR:021 Probe short error
)
Short-circuit detection can also be controlled by asserting the active-low PRB_CHECK EXT I/O signal.
Asserting the PRB_CHECK signal while measuring causes short-circuit detection to be performed after the
end of measurement (p. 85).
About Probe Short-Circuit Detection
• If probes are connected to the DUT during probe short-circuit detection, the short circuit is detected. Ensure
that the probes have sufficient time to separate from the measurement object.
• Probe short-circuit detection occurs within about 1 ms.
• The threshold for probe short-circuit detection is fixed at 500
, so if the resistance between CUR and POT
probes is larger, detection is not possible.
4.8
Test for Short-Circuited Probe
(Probe Short-Circuit Detection Function)
Foreign Object
POT
CU
R
DUT Electrode
DUT
Timing Chart
(Probe Short-Circuit Detection)
Probe short-circuit
PRB_SHORT
Comparator
Hi, IN, Lo
End of Measurement
EOM
Measurement
Start
TRIG
Contact
Transport
Measurement
stage
Available
Short-Circuit
Detection
Transport
DUT1 contact
DUT2 contact
Transport
Probe short-circuit
DUT1
measurement
DUT2
measurement
Short-Circuit
No Decision
Detection timing setting
•
Even while the probe short-circuit detection function is set to be disabled, short-circuit detection is performed when the
EXT I/O PRB_CHECK signal is asserted.
•
When the internal trigger [TRG: INT] source is selected, short-circuit detection is not performed after the end of mea-
surement. However, short-circuit detection can still be executed by asserting the PRB_CHECK signal.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......