3.3 Setting the Measurement Speed
29
3
The measurement speed can be set to FAST, MED (medium), or SLOW. A slower measurement
speed provides greater measurement precision, and a faster measurement speed results in greater
susceptibility to environmental noise. Ensure that measurement cables and the sample are suffi-
ciently shielded.
3.3
Setting the Measurement Speed
1
Open the Basic Settings screen.
2
Select the measurement speed.
3
Return to the Measurement screen.
Relationship Between Measurement Range and Speed
(factory defaults)
Measure-
ment Range
LOW POWER: OFF
LOW POWER: ON
FAST
MED
SLOW
FAST
MED
SLOW
100m
3.8 ms
13 ms
43 ms
36 ms
Integration time can be optionally set
for each range (p. 44).
Upper value:
for 50-Hz power line frequency
Lower value:
for 60-Hz power line frequency
Tolerance:
±
10%
±
0.2 ms
1000m
2.0 ms
6.4 ms
41 ms
35 ms
2.5 ms
12 ms
42 ms
35 ms
10
1.6 ms
6.0 ms
41 ms
34 ms
2.5 ms
12 ms
42 ms
35 ms
100
0.9 ms
3.6 ms
21 ms
17 ms
1.7 ms
6.1 ms
41 ms
34 ms
1000
0.9 ms
3.6 ms
21 ms
17 ms
7.2 ms
12 ms
47 ms
40 ms
10k
1.0 ms
3.6 ms
21 ms
17 ms
100k
1.3 ms
3.8 ms
21 ms
18 ms
1000k
2.5 ms
6.0 ms
21 ms
18 ms
10M
5.3 ms
23 ms
20 ms
23 ms
20 ms
100M
26 ms
22 ms
46 ms
39 ms
86 ms
72 ms
The Basic Settings screen appears.
1
Selection
2
See table below
(default)
Press the up/down cur-
sor keys to change the
setting.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......