
Chapter 10 Specifications
169
10
(5) OVC (Offset Voltage Compensation)
(6) Integration Time Setting
(7) Faulty Measurement Detection
Operation
Reverses measurement current polarity to eliminate offset voltage effects
Applicable ranges
LP OFF: 100 m
to 10
range
LP ON: All ranges
Operation
Sets the voltage detection acquisition time span (for each range)
Default state
Table 2. Integration Time Settings
Integration time is doubled when OVC is enabled.
Setting range
0.1 ms to 100.0 ms, per PLC
*
setting
1 to 5PLC: 50 Hz, 1 to 6PLC: 60 Hz
*
PLC: Power Line Cycle, One power line cycle (at 50 or 60 Hz)
Range
Low Power OFF
Low Power ON
Integration Time
OVC
Integration Time
OVC
FAST
MED
SLOW
FAST
MED
SLOW
100m
0.5 ms
5.0 ms
1PLC
ON
1000m
0.3 ms
2.5 ms
1PLC
ON
0.5 ms
5.0 ms
1PLC
ON
10
0.3 ms
2.5 ms
1PLC
ON
0.5 ms
5.0 ms
1PLC
ON
100
0.3 ms
3.0 ms
1PLC
OFF
0.3 ms
2.5 ms
1PLC
ON
1000
0.3 ms
3.0 ms
1PLC
OFF
0.3 ms
2.5 ms
1PLC
ON
10k
0.3 ms
3.0 ms
1PLC
OFF
100k
0.5 ms
3.0 ms
1PLC
OFF
1000k
1.5 ms
5.0 ms
1PLC
OFF
10M
2.5 ms
1PLC
1PLC
OFF
100M
1PLC
2PLC
4PLC
OFF
Out-of-Range Detection Function
Operation
Indicates under- or over-range values in the following conditions:
•
Measured value is outside of the measurement range
•
The relative calculation value is outside of the display range
•
Measured value is outside of the A/D converter input range
•
The zero-adjusted value is outside of the display range
Ex.: In the 1
range with 0.5
zero-adjustment in effect, measuring 0.1
pro-
vides a zero-adjusted value of -0.4
, which is outside of the display range.
Contact Check Function
Operation
Checks the connections between H
POT
and H
CUR
, and between L
POT
and L
CUR
ter-
minals (for each range)
Default state
ON, 200
Setting
ON/ OFF
Threshold setting
50
/ 100
/ 150
/ 200
/ 300
/ 400
/ 500
(reference value)
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......