APPENDIX 8
────────────────────────────────────────────────────
Appendix 3 Reference
────────────────────────────────────────────────────
Appendix 3.4 Recorder Function
1 2 3 4.......
.......100
1 DIV
Maximum voltage
Minimum voltage
One piece of sample data
Data obtained in the
set sampling period
1 2 3 4.......
.......100
1 DIV
Maximum voltage
Minimum voltage
One piece of sample data
[ Set a short sampling period ]
One division is equal to 100 samples.
One piece of sample data collected using the recorder function contains the
maximum and minimum voltage obtained in the set sampling period.
Therefore, this data is of a certain width.
When a short sampling period is set and the input waveform changes
slightly, a sudden disturbance such as noise will increase the difference the
between the maximum and minimum values. To eliminate this phenomenon,
set a long sampling period.