143
────────────────────────────────────────────────────
7.1 Numerical Calculation (MEM)
────────────────────────────────────────────────────
7.1.3 Making Settings for Numerical Evaluation
Flashing cursor
Function
display
Meaning
: Disable evaluation function.
: Return NG if result is outside specified
range.
Function
display
Meaning
: Stops at GO
: Stops at NG
: Stops regardless of the evaluation result
Depending on the results of the numerical calculation, a GO (pass) or NG
(fail) result is returned.
Evaluation criteria can be set independently for each of the calculation sets
No. 1 - No. 16.
Procedure
Screen: MEASUREMENT (STATUS)
1. Set each numerical calculation operation.
2. Move the flashing cursor to
COMP
, and use the
function key to select.
3. Make settings in the numerical value input
screen. (See 5.4 for details about the input
method.)
The range is -9.9999E+29 to +9.9999E+29.
It is not possible to set the upper limit under the
lower limit, or the lower limit over the upper
limit.
4. Set the evaluation interruption conditions.
Move the flashing cursor to Evaluation
Interruption Conditions and select using the
function keys.
Summary of Contents for MEMORY HiCORDER 8855
Page 2: ......
Page 20: ...xii Chapter Summary...
Page 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Page 28: ...8 1 2 Identification of Controls and Indicators...
Page 116: ...96 4 8 FFT Analysis Function...
Page 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Page 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Page 186: ...166 8 6 Moving Cursors to the Search Points...
Page 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Page 263: ......
Page 264: ......