7.2 Waveform Calculation (MEM)
────────────────────────────────────────────────────
7.2.1 Preparing for Waveform Processing
Function
display
Meaning
: Disable waveform processing.
: Enable waveform processing.
The various setting items are displayed.
: Execute Waveform Processing
NOTE
Function
display
Meaning
: Performs numerical calculation on all
waveforms
: Performs numerical calculation only on the
waveform between cursors A and B
Procedure
Screen: WAVE CALC (STATUS)
(1) Set the waveform calculation.
1. Press the
STATUS
key to display the waveform
calculation screen.
2. Move the flashing cursor to
Wave CALC
.
3. Use the function keys to make the selection.
The maximum recording length at which waveform processing calculation is
possible is 5000 divisions (20,000 divisions for 128 M words, 80,000
divisions for 512 M words). If the recording length exceeds this limit, the
waveform processing calculation is disable.
When the memory segmentation function or roll mode is used, waveform
processing is not possible.
When the memory split setting is turned off, you can browse the last 16
waveforms, but when calculations are applied to the waveform, all
waveforms except the block you are currently browsing will be lost. (All
waveforms except the block that calculations are applied to will be lost.)
The averaged waveform becomes available for waveform processing when
the averaging setting is turned OFF following measurement.
When scaling is set for the channel in which the processing result is to be
stored, scaling is not carried out and only the unit is valid. (See Section
5.3.)
When the waveform processing calculation executes simultaneously with
data acquisition, a forced termination displays the results being calculated.
In such cases, press the
F6
(
exec.
) key to reexecute the calculation.
(2) Setting the Meas Area
Allows you to specify the waveform calculation
range.
1. Move the flashing cursor to
Meas Area
.
2. Use the function key to specify the calculation
range.
Summary of Contents for MEMORY HiCORDER 8855
Page 2: ......
Page 20: ...xii Chapter Summary...
Page 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Page 28: ...8 1 2 Identification of Controls and Indicators...
Page 116: ...96 4 8 FFT Analysis Function...
Page 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Page 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Page 186: ...166 8 6 Moving Cursors to the Search Points...
Page 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Page 263: ......
Page 264: ......