![Hioki MEMORY HiCORDER 8855 Instruction Manual Download Page 139](http://html1.mh-extra.com/html/hioki/memory-hicorder-8855/memory-hicorder-8855_instruction-manual_2137316139.webp)
119
────────────────────────────────────────────────────
5.9 Setting the 8952 DC/RMS UNIT
────────────────────────────────────────────────────
Function
display
Meaning
: The response time is set to High Speed
(approximately 100 ms).
: The response time is set to High Speed
(approximately 800 ms).
: The response time is set to High Speed
(approximately 5 s).
(2) Setting the response
The response time can be selected from three levels,
High Speed, Standard, or Low Speed. Normally,
this selection is set to High Speed, but if the
frequency is low or fluctuations are frequent, more
stable display can be obtained by setting the
response time to Standard or Low Speed.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Response
item to
be set.
3. Use the function keys to make a setting.