APPENDIX 7
────────────────────────────────────────────────────
Appendix 3 Reference
────────────────────────────────────────────────────
Appendix 3.3 Measurement Limit Frequency
3
2
1
23
24
25
TIME/DIV
(s/DIV)
Sampling period
(s)
Measurement
limit frequency
(Hz)
5
µ
s/DIV
10
µ
s/DIV
20
µ
s/DIV
50
µ
s/DIV
100
µ
s/DIV
200
µ
s/DIV
500
µ
s/DIV
1 ms/DIV
2 ms/DIV
5 ms/DIV
10 ms/DIV
20 ms/DIV
50 ms/DIV
100 ms/DIV
200 ms/DIV
500 ms/DIV
1 s/DIV
2 s/DIV
5 s/DIV
10 s/DIV
30 s/DIV
1 min/DIV
2 min/DIV
5 min/DIV
50 ns
100 ns
200 ns
500 ns
1
µ
s
2
µ
s
5
µ
s
10
µ
s
20
µ
s
50
µ
s
100
µ
s
200
µ
s
500
µ
s
1 ms
2 ms
5 ms
10 ms
20 ms
50 ms
100 ms
300 ms
600 ms
1.2 s
3 s
800 kHz
400 kHz
200 kHz
80 kHz
40 kHz
20 kHz
8 kHz
4 kHz
2 kHz
800 Hz
400 Hz
200 Hz
80 Hz
40 Hz
20 Hz
8 Hz
4 Hz
2 Hz
0.8 Hz
0.4 Hz
0.13 Hz
0.067 Hz
0.033 Hz
0.013 Hz
As a general rule, to ensure that sampling catches the peaks of a typical sine
wave input on the display, more than 25 samples are required for each input
cycle.
The measurement limit frequency changes depending on the time axis range.
Summary of Contents for MEMORY HiCORDER 8855
Page 2: ......
Page 20: ...xii Chapter Summary...
Page 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Page 28: ...8 1 2 Identification of Controls and Indicators...
Page 116: ...96 4 8 FFT Analysis Function...
Page 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Page 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Page 186: ...166 8 6 Moving Cursors to the Search Points...
Page 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Page 263: ......
Page 264: ......