Appendix 10 Frequently Asked Questions
A
33
Appendix
How can I acquire data contin-
uously starting on 1/4/2012,
daily from 9:00 to 17:00, for
one month?
Set as follows:
Measurement Criteria
[Interval]
-
[Cont]
: Off, 8 hr 0 min 0 s
[Repeat]
: On
Timer
[Start]
: On, 12 Year, 4 Mon, 1 Day,
9 Hour 0 Min 0 Sec
[Stop]
: On, 12 Year, 5 Mon, 1 Day, 9
Hour 0 Min 0 Sec
[Interval]
: On, 1 Day, 0 Hour 0 Min 0
Sec
3.3, "Setting Configu-
ration 1 – Recording
Settings" (p.42)
4.4, "Starting and
Stopping Recording
by Timer" (p.76)
Is it possible to measure on 20
channels using two synchro-
nized HiLoggers?
Measurement start can be synchro-
nized using an external trigger.
Sampling cannot be synchronized.
9.5, "Synchronous
Measurements with
Multiple Instruments"
(p.135)
Question
Remedial Action
Reference page
Measuring
Question
Remedial Action
Reference page
What should I do about zero
offset when the input is short-
ed?
Execute zero adjustment to com-
pensate for offset when the inputs
are shorted.
2.5, "Zero Adjust-
ment" (p.35)
Even though a signal is only in-
put on CH1, why do similar
waveforms appear on unused
channels?
An open-circuit input terminal can be
affected by signals on other chan-
nels. By closing the input terminal
circuit, normal measurement is pos-
sible. If this is impractical, disable
unused measurement channels, or
short the + and - terminals together.
How to display numerical data
values?
Press
WAVE/DATA
repeatedly to
cycle through the Waveform (only),
Wa Numerical Values, and
Numerical Values (only) displays.
Numerical values can be displayed
in a large font on the enlarged nu-
merical display.
Numerical values at specific loca-
tions can be displayed using A/B
cursors.
1.4, "Screen Configu-
rations" (p.18)
"Displaying Cursor
Values" (p.84)
"Viewing Input Sig-
nals as Numerical
Values" (p.87)
What happens to pulse inte-
gration when there are more
than one million pulses?
The counter saturates at
1,073,741,823 counts. If you expect
to exceed this count, we recommend
measuring with the Integration Mode
(
[Count]
) set to Instantaneous
(
[Inst]
), and summing later with a
spreadsheet program.
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...