9.1 Connecting to the External Control Terminals
130
• The external control ground terminal is not isolated from the
Memory HiLogger’s chassis ground. Make certain that there will
be no potential difference between the external control ground
terminal and the ground of any connected device. Otherwise, the
Memory HiLogger or device could be damaged.
• To avoid electric shock, use the recommended wire type to con-
nect to the external control terminals, or otherwise ensure that
the wire used has adequate current-handling capacity and insu-
lation.
Terminal Connections
Required items:
Recommended cables:
Single strand diameter
0.65 mm (AWG22),
Multi-strand
0.32 mm
2
(AWG22)
Usable cables:
Single strand diameter
0.32 to 0.65 mm (AWG28 to 22),
Multi-strand 0.08 to 0.32 mm
2
(AWG28 to 22)
Strand diameter
0.12 mm or greater (per wire)
Standard insulation stripping length: 9 to 10 mm
Button operation specified tool: Flat-blade screwdriver
(shaft diameter
3 mm, tip width 2.6 mm)
Single strand
Multi-strand
10 mm
1
Push in the button on the connector with a flat-
blade screwdriver or other tool.
2
With the button held in, insert the cable into the
cable connection hole.
3
Release the button.
The cable is locked.
1
2
3
Top View
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......