Chapter 10 Specifications
140
Usable device
USB flash drive
USB flash drive
CF card
/
USB flash drive common functionality
Drive-to-drive copying
Files and folders can be copied between drives.
Measurement Input Section
Input terminals
M3 screw type terminal block (2 terminals per channel)
Pulse inputs use a special connector
Measurement object
• Voltage
• Thermocouple (K, J, E, T, N, R, S, B)
• Pulse count integration (cumulative and instantaneous), revolu-
tion variations
Measurement ranges and resolution
Temperature
measurement
input range
K: -200
C to 1350
C
J: -200
C to 1200
C
E: -200
C to 1000
C
T: -200
C to 400
C
N: -200
C to 1300
C
R: 0
C to 1700
C
S: 0
C to 1700
C
B: 400
C to 1800
C
(JIS C 1602-1995)
*1: Upper and lower limits depend on sensor input range
Measurement Object Range
Measurable Range
Max. Resolution
Voltage
100mV f.s.
-100 mV to +100 mV
5
V
1V f.s.
-1 V to +1 V
50
V
10V f.s.
-10 V to +10 V
500
V
20V f.s.
-20 V to +20 V
1 mV
100V f.s.
-60 V to +60 V
5 mV
1-5V
1V to 5V
500
V
Temperature
(thermocouple)
2000
C f.s.
-200
C to 1800
C *1
0.1
C
Pulse Integration
1,000M pulse f.s.
0 to 1,000M pulse
1 pulse
Revolution Variation
5,000/n r/s f.s.
0 to 5,000/n r/s
1/n r/s
n = pulses/revolution, from 1 to 1,000
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......