3.4 Setting Configuration 2 – Input Channel Settings
50
Configure these settings for each voltage measurement channel.
Voltage Measurement Settings
1
Select a channel (
CH 1
to
CH 10
) for the
[Channel]
setting contents.
2
To enable measurement on the channel,
select On to display the check mark (
).
3
Select
[Voltage]
for the
[Input]
(input type)
setting.
4
Select the appropriate measurement ampli-
tude range for the
[Range]
setting.
Setting options:
100mV
, 1V, 10V, 20V, 100V, 1-5V
(
default setting)
1
2
3
4
: On
: Off
Instrumentation and Control Device Measurements
For instrumentation control signal measurements, select
[1-5 V]
. This range corre-
sponds to the 4-20 mA output standard for instrumentation control. This setting automat-
ically sets 1- and 5-volt lower and upper limits, respectively, in the 10 V range.
Measurement accuracy is ±10 mV ±5 mV. To measure 4-20 mA current, connect a 250
shunt resistor b and – analog input terminals (p. 32).
Use the scaling function to scale measurements to any arbitrary unit value for display (p.
58).
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......