Appendix 1 Error Messages and Remedial Actions
A
3
Appendix
401
File processing error
An unexpected fault occurred while processing
a file on the CF card or USB flash drive. Turn the
Memory HiLogger off and back on.
402
Cannot load this file.
The format of the file is incompatible with the
Memory HiLogger, or the file is corrupt.
Refer to 6.1, "About Saving and Loading Data"
(p.91) for loadable file formats.
403
Insert a Media.
A CF card or USB flash drive is not present.
Insert storage media.
404
Directory full
No more files can be created because the direc-
tory is full. Check or perform the following:
1. System reset (p.125)
2. Verify that the removal storage is specified by
Hioki (p.93)
3. Reformat the removal storage (p.95)
If the message persists after performing the
above, either the removal storage or HiLogger
may be damaged. Contact your dealer or Hioki
representative.
405
Not enough capacity
The removal storage as insufficient space to
save the file. Delete files to increase free space,
or replace the removal storage.
406
Process interrupted
(not enough battery capacity).
There is inadequate battery life remaining.
Charge the battery or connect the AC adapter. If
you check the above and find no problem, the
CF card or instrument may be broken. Contact
your dealer or Hioki representative.
430
Some data was not saved.
The removal storage is not inserted, or does not
have enough free space.
Insert a removal storage and manually save
needed data.
431
Insert media to save waveform
data. Insert recording media.
Insert a removal storage.
432
Change media.
Removable storage cannot be accessed.
Replace the CF card or USB flash drive.
434
Media is full or cannot delete
oldest wave file.
This may appear while measuring using wave-
form auto save.
A file could not be saved: Not enough space on
the removable storage.
435
Media is nearly full.
Not enough space on the removal storage.
Replace with a new removal storage.
436
Saving intterruptted (low battery).
Battery charge is too low to save. Recharge, or
connect the AC adapter.
Message
Remedial Action
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......