3.4 Setting Configuration 2 – Input Channel Settings
51
3
3
Chapt
er
3 Sett
ing be
fore M
easure
ment
Configure these settings for each temperature measurement channel.
Temperature Measurement Settings
1
Select a channel (
CH 1
to
CH 10
) for the
[Channel]
setting contents.
2
To enable measurement on the channel,
select On to display the check mark (
).
3
Select
[Tc]
for the
[Input]
(input type) set-
ting.
4
Select the type of thermocouple to be
used.
Setting options:
K
, J, E, T, N, R, S, B
(
default setting)
5
Enable or disable
[Burn Out]
.
Setting options:(
default setting)
6
Select the reference junction compensa-
tion (RJC) method for the
[RJC]
setting.
Setting options:(
default setting)
Off
Do not detect broken (disconnected)
wires.
Normally when the thermocouple is dis-
connected, the value will fluctuate, etc.
and will not be stable.
On
Detect broken wires.
(This setting is available only when the re-
cording interval is at least 100 ms, and the
50 Hz or 60 Hz digital filter is enabled.)
When the wires are disconnected,
“OVER” will be displayed as the measure-
ment value.
Int
Use the Memory HiLogger’s internal com-
pensation. Measurement accuracy is the
sum of the temperature measurement ac-
curacy and the RJC accuracy.
Ext
Use an external RJC method instead of
the Memory HiLogger’s internal compen-
sation. Memory HiLogger measurement
accuracy is equivalent to the temperature
measurement accuracy.
1 2
: On
: Off
Temperature Measurement
Range
(Depends on thermocouple type)
JIS C 1602-1995
With type B, the display shows 0
to 400°C, but accuracy is not
guaranteed.
K
-200°C to 1350°C
J
-200°C to 1200°C
E
-200°C to 1000°C
N
-200°C to 1300°C
T
-200°C to 400°C
R
0°C to 1700°C
S
0°C to 1700°C
B
400°C to 1800°C
3 4
5
6
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......