Appendix 8 Concerning Noise Countermeasures
A
15
Appendix
The Importance of Device Grounding
Ground the instruments (Model LR8431-20) securely
The LR8431-20 is designed so that the chassis is grounded when the three-prong
power cord from the adapter is plugged directly into a grounded, three-prong recep-
tacle.
If a three-prong-to-two-prong adapter must be used on the AC adapter power cord
and therefore the instrument is not grounded, the instrument can still be grounded by
connecting a grounded wire to the GND terminal (a pushbutton terminal) located on
the side of the LR8431-20.
Securely ground the chassis of the device being tested.
Securely connect the chassis of the device being tested to a good ground.
Connecting the chassis GND on the signal side to the instrument chas-
sis
When connecting the chassis ground of the LR8431-20 to the chassis ground of the
device being tested and then to ground, use a wire that is as short and as thick as
possible to bring both pieces of device to equal potential.
Thermocouple
Device being
Tested
Instrument
Grounding Both the Device Being Tested
and the Instrument
Chassis GND of In-
strument being Test-
ed
GND of the
Instrument
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......