9.5 Synchronous Measurements with Multiple Instruments
135
3
9
Chapt
er
9 Externa
l Control
Multiple instruments can be synchronized using the external control terminals.
Parallel synchronization
Set one instrument to master, and set the others
to slave.
9.5 Synchronous Measurements with
Multiple Instruments
GND
EXT.TRIG
TRIG OUT
GND
Connection example
EXT.TRIG
TRIG OUT
GND
EXT.TRIG
TRIG OUT
GND
Master
Slave
Slave
Set the master instrument.
1
Press the
SET
key to open the [System]
screen.
2
Set the
[Trig Out]
setting to
[Start]
.
Set the slave instrument.
1
Press the
SET
key to open the
[System]
screen.
2
Set the
[
External Trg In
]
setting to
[
Trigger
]
.
3
Press the
SET
key to open the [Trig&Alm]
screen.
4
Set as follows.
Trig Func: On
Timing: Start
Trig Source Start: OR
Ext Trig Start: On
Turn the start trigger setting for all the channels
(analog, pulse) to
Off
. Failure to do so may start the
measurement without synchronization.
Press the
START/STOP
key of the slave instrument to start the measurement (trig-
ger waiting state). In this state, press the
START/STOP
key of the master instru-
ment to start the measurement. The slave instrument will start the measurement at
the same time.
1,2
3,4
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......