9.1 Connecting to the External Control Terminals
129
3
9
Chapt
er
9 Externa
l Control
The external control terminals on the Memory HiLogger support trigger signal input
and output.
External
Control
Chapter 9
ALARM (Output)
A signal is output whenever the alarm
criteria are satisfied (p. 134).
EXT.TRIG (External trigger input)
Input external signals (p. 131).
Top View
GND
(common with instrument ground)
TRIG.OUT (Trigger output)
Output a signal when a trigger
event occurs (p. 132).
9.1 Connecting to the External Control
Terminals
To avoid electrical hazards and damage to the instrument, do
not apply voltage exceeding the rated maximum to the exter-
nal control terminals.
I/O terminals Maximum input voltage
Input
EXT.TRIG
0 to 7 V DC
Output
TRIG.OUT
30 V DC
5 mA max, 200 mW max
ALARM
To prevent electric shock accidents and damage to the equip-
ment, always observe the following precautions when making
connections to external terminal blocks and external connec-
tors.
• Before making connections, turn off the power on the
instrument and the equipment to connect.
• Do not exceed the specified signal levels for signals sup-
plied to external terminal blocks.
• Ensure that devices and systems to be connected to the
external control terminals are properly isolated.
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......