4.2 Alarm Output
74
1
For the
[Channel]
, select a channel for which you
want to set alarm criteria.
2
Select
[Alarm on.]
to enable alarm criteria settings.
3
Select the alarm criteria type from the
[Cond]
setting options.
Setting options:(
default setting)
Alarm criteria settings are not available unless a check
mark [
] is displayed next to the channel number.
4
Set the alarm criteria.
If you have selected
[Level]
.
Setting options:
If you have selected
[Window]
.
Setting options:
Off*
Disable alarms for this channel.
Level
Trigger an alarm when the input reaches a
specified level.
Window
Trigger an alarm when the input transitions
through upper and lower limits (window).
Alarm output occurs when the input signal rises
above the specified level threshold.
Alarm output occurs when the input signal falls be-
low the specified level threshold.
In
Alarm output occurs when the input signal enters
the range defined by the thresholds.
Out
Alarm output occurs when the input signal exits
the range defined by the thresholds.
1
Setting Alarm Criteria for Each Channel
ALM appears when no
channel output is checked
(
). In this case, select
[Alarm on.] to enable
alarm output settings, and
then set alarm criteria.
Select how you want to combine alarm output crite-
ria from the
[Alarm]
setting options.
Setting options:(
default setting)
OR*
Alarm output occurs when the alarm criteria specified
for any alarm-enabled channel are satisfied.
AND
Alarm output occurs only when the alarm criteria for
every alarm-enabled channel are satisfied.
Selecting Combined Alarm Output Criteria
Select the [Trig & Alm] Screen.
If set to [Off], change it to
[On].
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......