6.4 Loading Data on the Memory HiLogger
103
3
6
Chapte
r 6
Sa
ving & Loading Dat
a
Previously stored binary waveform data, captured screen images and saved
setting configurations can be reloaded into the Memory HiLogger (p.91)
Setting configurations saved in the Memory HiLogger’s memory or on a CF
card or USB flash drive can be reloaded.
See:
"Saving Setting Configurations" (p. 102)
6.4 Loading Data on the Memory HiLogger
Loading a Setting Configuration
Select the
[Setting]
Screen.
Select
[Load Set.]
.
1
2
Apply
Select the configuration
file to load.
(CONF
nnnn
.SET)
Memory HiLogger settings are reconfigured to those in the loaded configuration file.
Apply
The loading dialog appears.
3
Select whether to load a set-
ting configuration from
LR8431-20’s memory or CF
card
or USB flash drive.
Apply
(If you selected
[CF Card/USB
flash drive]
)
When Loading from CF card or USB flash drive
Configuration files stored in the [HIOKI_LR8431]-
[CONFIG] folder on the CF card or USB flash drive
are displayed.
When Loading from Memory HiLogger Memory
The list of configurations stored in Memory
HiLogger memory is displayed.
4
(If you selected
[Memory]
)
Select
[Load]
for the con-
figuration you want to load.
Apply
To automatically load settings data (auto-setup)
Settings data saved under the filename “STARTUP.SET” in the “HIOKI_LR8431”-”CONFIG”
folder will be loaded automatically when the instrument is turned on. If settings have been saved
on both the CF card and USB flash drive, the CF card’s settings will take precedence. (“STAR-
TUP.SET” may not be loaded from USB flash drives that take a long time to be recognized.)
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......