86
Table 7.3 Setting Control Program Messages 5/13
Mnemonic Description
Format
ELC
Electrode data setting
d1 (SIZE/ACTL selection)
(0: ACTUAL, 1:SIZE)
d2 (Electrode inside diameter SIZE)
(0.0 ~ 999.9mm)
d3 (Electrode outside diameter SIZE)
(0.1 ~ 1199.9mm)
d4 (Sample thickness SIZE)
(0.001 ~ 30.000mm)
d5 (Random coefficient ACTUAL)
(0.01 ~ 999.99mm)
[Note] Be sure to achieve (Electrode inside
diameter < Electrode outside diameter).
Ignored if not complied with and only
settable parts are set.
Format]
ELC d1 d2 d3 d4 d5
d1: NR1 format
d2: NR2 format
d3: NR2 format
d4: NR2 format
d5: NR2 format
ELC?
Query response electrode data is same as
setting.
[Format]
ELC?
[Response]
d1, d2, d3, d4, d5
CNF
Operation environment setting
d1 (Interlock control provided, not provided: 0 ~
1)
0: CONNECT 1: CUTOFF
d2 (Beep tone provided, not provided: 0 ~ 1)
0: OFF 1: ON
d3 (Beep tone good/no good: 0 ~ 1)
0: NO 1: GO
d4 (Click tone provided, not provided: 0 ~ 1)
0: OFF 1: ON
[Format]
CNF d1, d2, d3, d4, d5
d1: NR1 format
d2: NR1 format
d3: NR1 format
d4: NR1 format
d5: NR1 format
CNF?
Query response of operation environment is
same as setting.
[Format]
CNF?
[Response]
d1, d2, d3, d4, d5
Summary of Contents for DSM-8542
Page 2: ......
Page 21: ...11 3 2 Block Diagram of DSM 8542...
Page 89: ...79 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 90: ...80 Command program header Fig 7 2 Program Header Syntax Chart...
Page 91: ...81 Fig 7 3 Syntax Chart of Data Part...
Page 138: ...128 11 Product Full View...
Page 139: ......
Page 140: ......