Table 7.3 Setting Control Program Messages 3/13
Mnemonic Description
Format
AVE
Averaging setting
d1 (Selection: 0 ~ 1)
0: OFF (Not done)
1: ON (Done)
[Format]
AVE d1
d1: NR1 format
AVE?
Query response in averaging is same as setting.
[Format]
AVE?
[Response]
d1
FIG
Setting number of effective columns for
measured value display
d1 (Effective columns: 2 ~ 5)
[Format]
FIG d1
d1: NR1 format
FIG?
Query response on setting number of effective
columns for measured value display is same as
setting.
[Format]
FIG?
[Response]
d1
RNG
Current range setting
AUTO/HOLD selection and range setting during
HOLD
d1 (Selection: 0 ~ 1)
0: HOLD
1: AUTO
d2 (Range during HOLD: 1 ~ 8)
0: 10
μ
F 4:
1nF
1: 1
μ
F
5: 100pF
2: 100nF
6: 10pF
3: 10nF
7: 1pF
Capacity of integrating capacity is set in range
changeover
[Format]
RNG d1, d2
d1: NR1 format
d2: NR1 format
RNG?
Query response on current range is same as
setting.
[Format]
RNG?
[Response]
d1, d2