40
(1) Press the [PROG] key on the Measurement screen.
The following Sequence Program Creation screen will be displayed.
M A K E S E Q U E N C E P R O G R A M
P R O G R A M N o . : [ 0 ]
D I S C H G 1 : [ 0 . 0 ] S E C
C H A R G E : [ 0 . 0 ] S E C
M E A S T I M E : [ 0 . 1 ] S E C
D I S C H G 2
: [ 0 . 0 ] S E C
D O W N U P
Fig. 5.8 Sequence Program Creation Screen
The Function keys [F2] and [F3] will respectively change to the DOWN and UP functions.
A preset value can be chosen keying the DOWN and UP keys in addition to directly
inputting it through the numeric input keys.
After finishing setting, return to the Measurement screen by pressing the [ENTER] key.
(2) Program No. setting
Input in “PROGRAM NO.” a program No. for a program to be created.
Ten types of sequence programs, 0 to 9, can be created.
(3) Charging time, measuring time and discharging time setting
In sequence program measurement, the operational sequence of DISCHG1 --
CHARG -- MEAS TIME -- DISCHG2 cannot be changed.
(a) Setting discharging time before measurement (DISCHG1)
Set a time to discharge the sample before starting measurement.
0 to 999.9s can be set.
(b) Setting charging time before measurement (CHARGE)
Set a time to charge the sample by impressing measuring voltage.
0 to 999.9s can be set.
(c) Setting measuring time (MEAS TIME)
Set a time to measure a sample.
0 to 999.9s can be set.
(d) Setting discharging time after measurement (DISCHG2)
Set a time to discharge the sample after measurement.
0 to 999.9s can be set.
Summary of Contents for DSM-8542
Page 2: ......
Page 21: ...11 3 2 Block Diagram of DSM 8542...
Page 89: ...79 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 90: ...80 Command program header Fig 7 2 Program Header Syntax Chart...
Page 91: ...81 Fig 7 3 Syntax Chart of Data Part...
Page 138: ...128 11 Product Full View...
Page 139: ......
Page 140: ......