46
(4) Setting power supply noise filter
Set the noise filter in “FILTER: OFF/ON.”
OFF: The filter is not used (High-speed changeover mode: Fast)
ON: The filter is used Low noise mode: Slow)
Standard setting
Power Supply Noise Filter
Noise of measuring voltage output greatly affects the measurement accuracy in
measurement of samples which have a high electrostatic capacity such as
capacitors. For this reason, the meter is equipped with a filter to reduce noise
of measuring voltage output.
In regular measurement, stable measurement is possible by turning the filter on.
The condition in which the filter is turned on is called the “low-noise mode.”
The low-noise mode enables stable measurement with less noise. However,
one disadvantage with this mode is that the changeover speed of measuring
voltage becomes slow depending on the filter response time.
Select the high-speed changeover mode with the filter turned off in
measurement that requires fast changeover of measuring voltage.
The low-noise mode with the filter turned on is set during preshipment
inspection of the meter at the HIOKI factory
Table 5.8
Current Limit Values (Example when PSU-8541 is used)
Set Current Limit Value
Current Value
Voltage
Range
Measuring
System
Charging
System
Measuring
System
Charging
System
Total
Load
Voltage
Source
Current
Capacity
0.1~1000 5mA 5mA 5mA×4
5mA×5×4
120mA
600mA
5mA 10mA
5mA×4
10mA×5×4
220mA
(200mA)
5mA 25mA
5mA×4
25mA×5×4
520mA
5mA 50mA
5mA×4
50mA×2×4
420mA
10mA 5mA 10mA×4
5mA×5×4 140mA
10mA 10mA 10mA×4
10mA×5×4
240mA
10mA 25mA 10mA×4
25mA×5×4
540mA
10mA 50mA 10mA×4
50mA×2×4
440mA
25mA 5mA 25mA×4
5mA×5×4 200mA
25mA 10mA 25mA×4
10mA×5×4
300mA
25mA 25mA 25mA×4
25mA×5×4
600mA
25mA 50mA 25mA×4
50mA×2×4
500mA
50mA 5mA 50mA×4
5mA×5×4 300mA
50mA 10mA 50mA×4
10mA×5×4
400mA
50mA 25mA 50mA×4
25mA×4×4 600mA
50mA 50mA 50mA×4
50mA×2×4 600mA
* When using the charging terminal, make sure that total current does not exceed the
capacity.
* A voltage drop will result with the voltage source if the total load current exceeds the
current capacity of the voltage source.
* Continuous rating of current capacity is specified in parentheses ( ).
* A voltage drop will result with the voltage source if the rating is exceeded by a
continuous load
Summary of Contents for DSM-8542
Page 2: ......
Page 21: ...11 3 2 Block Diagram of DSM 8542...
Page 89: ...79 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 90: ...80 Command program header Fig 7 2 Program Header Syntax Chart...
Page 91: ...81 Fig 7 3 Syntax Chart of Data Part...
Page 138: ...128 11 Product Full View...
Page 139: ......
Page 140: ......