P/N 53063-30, Manual revision: B
A-13
Providing Reference Information
E
XPLAINING
IEC 60601-1-2:2001
STANDARD
REQUIREMENTS
Electromagnetic immunity
The Cell Saver 5+ system is intended for use in the electromagnetic environ-
ment specified in the following tables and the operator must ensure that
each system is used in such an environment.
IEC 60601-1-2:2001, Table 201: Guidance and manufacturer’s declaration - electromagnetic
immunity
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CSIPR 11
Group 1
The CS5+ system uses RF energy only for its in-
ternal functions. Therefore, its RF emissions
are very low and are not likely to cause any in-
terference in nearby electronic equipment.
RF emissions
CSIPR 11
Class A
The CS5+ system is suitable for use in all es-
tablishments, other than domestic establish-
ments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies