P/N 53063-30, Manual revision: B
7-5
Sequestering using the Cell Saver 5+ System
Setting the processing parameters
Performing sequestering is a
MANUAL
operation requiring operator inter-
vention to move from one stage to another. The process should be
completely understood before the sequestering protocol is used.
The preset parameters for both sequestering methods are set at:
!
Pump draw rate: 60 ml/min.
!
Centrifuge speed: 4750 rpm.
!
Empty rate: 200 ml/min.
To change the settings the operator will need to enter the sequester setup
menu from the STANDBY state screen display.
"
Press the Modify key to receive the following information:
Figure 7-4, Example of the Sequester setup menu
"
Press the Select key to scroll the list and highlight the parameter.
"
Press the Arrow keys [
!"#
] to adjust the value displayed on the screen.
"
Press the Modify key to return to the STANDBY state screen display.
Before starting the sequestering process, the operator should verify that the
waste bag clamp is open and the collection bag clamps are closed. While
the bowl is filling, the operator should ensure that the air in the disposable
system is entering the waste bag. The air will be removed from waste bag
during the EMPTY state.
Warning: Only one (1) unit of blood should be collected and processed at a
time. Unless directed by a physician, the second unit of blood should not
be drawn from the patient until the first unit of packed autologous red
blood cells has been returned.