P/N 53063-30, Manual revision: B
5-17
Cell Saving using Automatic Operation
Note: The NO WASH option may be permanently hidden from the Cell
Saver SETUP screen by Haemonetics field service if the physician wants to
prevent erroneous use of this option by the Cell Saver 5+ operator.
Reservoir Level and Resume at Level
options
These options allow the operator to customize the thresholds used by the
reservoir level sensor to initiate or resume the FILL state from the STANDBY
state:
!
Reservoir level
(initial): target volume in the reservoir required to
automatically initiate a fill cycle when the bowl is empty.
!
Resume at level
(resume): target volume in the reservoir to required to
automatically resume a fill cycle when the bowl is partially filled.
To customize the settings for the reservoir sensor level:
"
Press the Modify key to access the Cell Saver setup display.
"
Press the Select key until either
Reservoir level
or
Resume at Level
is
highlighted.
"
Press the Arrow keys to modify the values.
"
Press the Modify key to return to the STANDBY screen display.
Level Sensor option
To disable the reservoir level sensor function in the STANDBY state, the op-
erator can set the
Level Sensor
option to OFF. By doing this, only pressing
the Start key will initiate the FILL state from the STANDBY state.
This function can be modified as follows:
"
Press the Modify key to access the Cell Saver setup display.
"
Press the Select key until
Level Sensor
is highlighted.
"
Use the Arrow keys to select OFF.
AutoCycle option
This option allows the operator to start a new FILL cycle automatically, ime-
diately after completing an EMPTY cycle, without the level sensor and when
the collection device is mechanically not compatible with the level sensor.
To enable this feature, the operator should set the
AutoCycle
option to ON.
Note: The ability to use this option must be enabled by Haemonetics field
service.
!
When this parameter is set at OFF, the FILL cycle can be started
through level sensor detection or by pressing the Start key.
!
When this parameter is set at ON, after an EMPTY cycle, the FILL cy-
cle will start automatically from the STANDBY state after 3 seconds of
pause time. The whole procedure cycle: FILL, WASH, EMPTY,
STANDBY, FILL, ... will continue to be performed, as long as trip to
WASH is automatic, and there is blood in the collection container.