P/N 53063-30, Manual revision: B
5-18
Cell Saving using Automatic Operation
Speed regulation option
As previously discussed, the pump speed in the FILL, CONC and WASH
states is optimized according to the blood quality determined by the effluent
line sensor. To disable this feature, the operator should set the
Speed regula-
tion
option to OFF.
When this parameter is set at OFF, the pump speed selected by the operator
during the FILL, WASH and CONC states will be used systematically for all
subsequent cycles unless another bowl type is detected.
One application for use is during emergency procedures where processing
speed is more critical than optimized quality.
When pump speed regulation is disabled, the following message will be dis-
played during the FILL and WASH states to remind the operator that the line
sensor control of the pump speed is not active: