P/N 53063-30, Manual revision: B
4-8
Preparing for a Cell Saver 5+ Procedure
I
NSTALLING
A
C
ELL
S
AVER
DISPOSABLE
SET
At this point the operator can press the Help key for abbreviated installation
instructions, as illustrated in
.
The following para-
graphs provide further details related to the HELP installation list.
Note: The setup instructions provided by the Help key are merely reminders
and are not intended to serve as a substitute for formal Haemonetics training
on the use of the CS5+ device.
Preparing the collection system
"
Install a collection reservoir on the IV pole and clamp the reservoir
drain.
"
Open the Aspiration and Anticoagulation set using aseptic technique.
"
Pass the sterile inner wrapped line onto the sterile field and in the
sterile field, attach a plastic suction wand.
"
Connect the Aspiration and Anticoagulation set to the reservoir.
Attach wall suction to reservoir and
set suction at a minimal accept-
able level (recommended –80 to –150 mmHg)
. Close the roller
clamp on the anticoagulant line.
Figure 4-6, Connecting the A&A assembly to the reservoir
Warning: When using a vacuum source, the operator should be aware that
vacuum force greater than –200 mmHg may cause hemolysis.
1. Connector to vac-
uum source (yellow
cap)
2. Reservoir
3. Connector to aspi-
ration line (blue
cap)
1.
3.
2.