P/N 53063-30, Manual revision: B
A-10
Providing Reference Information
43
Pump error.
Clean pump rotor.
Check tubing positioning.
Pump stalled error.
Pump rotation is blocked.
44
Pump error.
Incorrect pump speed detected.
Incorrect pump speed.
Pump speed target could not be reached.
*45
Please be sure pump platen and valve
door are closed.
Pump lever or manifold latch error.
The pump lever or manifold latch is/are
not locked.
46
Sensor is not calibrated.
Reservoir level.
Level sensor calibration error.
Level sensor calibration is out of the oper-
ating range.
47
Sensor is not calibrated.
Waste bag weigher.
Waste bag weigher calibration error.
Waste bag weigher calibration is out of
the operating range.
*49
Unexpected sensor reading.
Line sensor out of range.
Clean line sensor and check disposable
insertion.
Line sensor tubing installation error.
Line sensor signal with tubing installed is
below specification.
Note: The asterisk (*) signifies that the specified error code is not recorded.
Code
Error message on the screen
Description