A variety of disposable sets are available for use with the CS5+ device. This
section provides a general understanding of the various elements, uses and
operating characteristics. Detailed installation instructions and further speci-
fications will be provided in subsequent chapters. The following figure illus-
trates a typical disposable set designed for a CS5+ procedure:
Figure 2-11, Example of a Cell Saver 5+ disposable set
Harness tubing
The harness consists of three lines:
!
The yellow-coded tubing connected to the saline solution.
!
The red-coded tubing connected to the unprocessed blood source.
!
The blue-coded tubing connected to the reinfusion bag.
All three color-coded lines pass through the pinch valves on the device and
are joined at a tubing manifold. A clamped line sensor is located under the
blue line to warn of an occluded line to the reinfusion bag.
After the junction, the single tubing line passes first through the pump, then
the air detector. After passing through the air detector, the single line enters
the centrifuge well through a slot in the rim where it is held down by the
cover. Once inside the well, the line is attached to the inlet port of the bowl.
The effluent line, attached to the outlet port of the bowl, exits the centrifuge
through the effluent line sensor and is then connected to the waste bag.
1. Centrifuge bowl
2. Tubing manifold
3. Red line
4. Collection reser-
voir connector
5. Yellow line
6. Saline bag spikes
7. Blue line
8. Reinfusion bag
9. Effluent line
10. Waste bag