P/N 53063-30, Manual revision: B
2-8
Describing the Cell Saver 5+ System Components
Process state keys
Figure 2-7, Process state key section
These keys can be used when backlit at specific points during the automatic
mode; they will be permanently lit for use during the manual mode. Use
will vary according to the selected CS5+ protocol as described in the
following table:
Key
Purpose
Fill
Cell Saver:
Used to pump fluid from the reservoir into a
spinning bowl.
Sequester:
Used to pump whole blood from a blood source
into a spinning bowl.
Conc
Cell Saver / Sequester:
Used to pump fluid from the product
bag into a spinning bowl. Supplements “Fill” and is usually
selected when the bowl is partially full and the reservoir is
empty.
Wash
Cell Saver:
Used to pump saline into a spinning bowl from
the saline solution bag.
Sequester:
Not used.
Empty
Cell Saver / Sequester:
Used to pump fluid from a stationary
bowl into a product bag.
Return
Cell Saver:
Used to pump fluid from a stationary bowl back
into the reservoir or into a bypass circuit. Can be used as an
alternative to the
Empty
key where the contents of the bowl
are pumped back (returned) through the “Fill” line instead of
being sent (emptied) to the product bag.
Sequester:
Not used.