P/N 53063-30, Manual revision: B
A-16
Providing Reference Information
The CS5+ system is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of
the CS5+ system can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications
equipment (transmitters) and the CS5+ system as recommended below,
according to the maximum output power of the communications equipment.
IEC 60601-1-2:2001, Table 206: Recommended separation distance between portable RF
communications equipment and the CS5+ device
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz, the separation for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
d
3 5
,
V
1
---------
P
=
d
3 5
,
E
1
---------
P
=
d
7
E
1
-----
P
=