P/N 53063-30, Manual revision: B
4-15
Preparing for a Cell Saver 5+ Procedure
Setting up the saline solution
"
Hang the saline wash solution bags on the lower pigtail of the IV pole
so that they hang in a cascading manner.
"
Close the clamps on both of the yellow-coded wash lines.
"
Spike the saline bags and unclamp the lines.
Note: Each wash cycle requires a volume of saline solution which will
depend on the size of the bowl in use.
!
Latham bowl 225 ml: 1000 ml saline solution.
!
Latham bowl 125 ml: 750 ml saline solution.
!
70mL Bowl: 300 ml saline solution.
Inspecting the installation
"
Inspect all parts of the disposable set and verify that there are no
twists, kinks or flat spots.
"
Verify that all connections are secure and all appropriate clamps and
covers are closed.