P/N 53063-30, Manual revision: B
5-6
Cell Saving using Automatic Operation
Emptying the bowl
When the appropriate volume of saline has entered the bowl, the CS5+ sys-
tem will automatically stop the pump. The centrifuge will stop spinning and
the red blood cells suspended in saline solution will be pumped from the
bowl into the reinfusion bag. The following information will be displayed
during the EMPTY state:
Figure 5-5, Example of an EMPTY state screen display
After one bowl has been processed, the system will start processing again
when the preset reservoir level is detected. If there is no further blood to pro-
cess, the operator should power off the device and remove the disposable
set.
Monitoring the waste bag
During the procedure, the CS5+ system will monitor the amount of fluid
collected in the waste bag and alert the operator to change or drain the con-
tents of the waste bag when it is almost full. If the waste bag does become
full, the CS5+ system will produce an audible alarm, move into a PAUSE
state and indicate that the waste bag is full. Once corrective action is taken,
the operator should:
"
Press the Start key to resume processing.
Caution: A full waste bag should be changed or emptied only when the
bowl is emptied of blood (and filled with air). The waste bag may be par-
tially emptied through the drainage port at any time as long as the fluid level
in the bag does not fall below the waste bag drainage port.