P/N 53063-30, Manual revision: B
A-7
Providing Reference Information
D
ESCRIBING
C
ELL
S
AVER
5+
ERROR
CODES
The CS5+ data acquisition will automaticaly store all critical errors occur-
ring during a procedure using codes.
The following table lists these error codes with a brief description:
Code
Error message on the screen
Description
1
System error.
Software timing error.
Software timing error.
2
Clamp error.
Check reservoir (red) line.
Remove tubing and inspect area around
clamp.
RED clamp position error.
Occurs whenever the clamp should be
open/closed but is not detected as open/
closed.
3
Clamp error.
Saline (yellow) line.
Remove tubing and inspect area
around
clamp.
YELLOW clamp position error.
Occurs whenever the clamp should be
open/closed but is not detected as open/
closed.
4
Clamp error.
Product (blue) line.
Remove tubing and inspect area around
clamp.
BLUE clamp position error.
Occurs whenever the clamp should be
open/closed but is not detected as open/
closed.
5
System error.
Pump driver error.
Pump error.
Software initialization error.
6
Centrifuge error.
Centrifuge unable to maintain
requested speed.
Check bowl for proper seating.
Illegal centrifuge speed.
Centrifuge speed request is above speci-
fied limit.
7
System error.
Keypad error.
Keypad error.
9
System error.
Processor error.
System fault, processor error.
*10
Reservoir level.
Check reservoir level sensor
and disposable.
Check tubing placement.
Check for occlusions in harness
and bowl.
Air detected too early in FILL mode.
Occurs whenever air is detected in FILL
mode before at least 50 ml have been
processed.
Probable cause: level sensor initiated a fill
cycle without fluid in the reservoir (accu-
racy problem).
11
System error.
Software stack error.
Stack empty error, software state
management error.
13
Solenoid error.
Solenoid task error.