installation & commissioning
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 59 of 213
[6.F]
Installation, Overview, Checklist
The authorized technician (distributor or service partner) has to check whether the installation of the device has been
performed properly, before the customer is introduced into the operation of the h/p/cosmos device. For detailed
information and instructions for installation and commissioning please refer to all respective individual chapters in this
manual. Among others, following points save to be performed or controlled:
No. Illustration
Description
[01]
Compare the delivery note with the installed running machine.
Are all parts like running machine, accessories, service box and
running machine folder complete and without damage.
[02]
Adjust the levelling sockets.
[03]
Tighten all fixation screws of the handrails.
[04]
Check the wall socket: Optical check for damages, burned
contacts, proper fixation and good ground contact (no paint, dirt,
deformation)
Check out the correct pin assignment of the wall socket with a
socket-tester with direct indication (see picture left).
As testing-device we recommend the Testavit Schuki 3 –
[cos15900] - Separate instruction available at h/p/cosmos.
Check the electrical connection of the device: Direct connection to
the wall socket with separate fuse, no danger of stumbling, no
extension cable, no multiple plug socket, separate circuit for the
treadmill. If a correct connection to the wall socket is not possible,
note this on the delivery note, inform the customer and put the
device out of operation (if necessary) and safe
against start-up.
accomplished
accomplished
accomplished