accessories & options
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 180 of 213
The running belt rotation reverse will be announced by an acoustic signal and the display
=
"r." Speed will be
indicated as a negative value while in operation
=
„-0.0 ... -5.0“.
[12.M1]
Pull Cord Actuator (l magnet)
This stop function is a safety feature that is
used with a waist belt. Especially with
reverse belt rotation (downhill) it is also
important to connect. The athlete or patient
is connected to the magnetic switch at the
UserTerminal via a ripcord at the waist belt.
In case of an emergency the magnet gets
ripped of from the UserTerminal and the
running machine stops immediately. The
ripcord is not a prevention of falling. At
removed magnet „PULL STOP“ will occur in
the display.
Safety harnesses such as safety stop cord type (safety lanyard with pull-cord actuator) and/or safety arch with
harness, chest belt and stop function and/or unweighting systems with patient vest and harness shall be adjusted
in a way, that the subject can not use the rear 30% of the running deck length.
Pull Cord Actuator (l magnet) with STOP function: [cos00098050043]
Chest-harnesses, patient-vests, waist-belts with leash and cuffs are not designed for direct
skin or mucous membrane contact. Wear appropriate clothing.
picture
type of stop
function
interface
communication
how to restart
Pull-cord
actuator (lanyard
with
magnetic switch
on the
UserTerminal and
waist belt on the
subject)
N
magnet is pulled off through
the cord
N
running belt & inclination
system stop fast
N
machine is NOT cut off from
main power of drive motor
and inclination system
N
display active and shows
“PULL STOP”
N
interface active, but no start,
speed or elevation signal
executed
N
keyboard inactive
communication
NOT interrupted
during
“pull stop” stage
N
reconnect magnet
to its position on
UserTerminal
N
press START key,
N
or new command
through interface