optional settings
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 102 of 213
[8.A2]
OP20
RS232 interface protocol: COM 1
Various ECG and Ergospirometry
systems support the h/p/cosmos coscom
protocol. Individual adaptation of
protocol on request. For further
information about the h/p/cosmos
coscom protocol see www.coscom.org
We recommend to use the h/p/cosmos
coscom protocol, as it provides a high
safety and functionality standard.
h/p/cosmos para control
®
and
h/p/cosmos para graphics
®
exclusively
work with the h/p/cosmos coscom
protocol.
For advanced h/p/cosmos coscom v3
connections with Baudrate 115200
please select OP20=20 . Please note
that the connected peripherie/software
must be approved for h/p/cosmos
coscom v3 with Baudrate 115200 (for
example h/p/cosmos para control 4.0).
§
indicates flashing: number of the RS232 interface protocol
$
and
%
reports:
OFF = RS232 not active / no protocol / interface deactivated
1 = h/p/cosmos coscom v1, v2, v3 with Baudrate 9600 bps
( standard setting COM 1 and COM 2)
2 = JAEGER OXYCON (up to 1995); if available use: h/p/cosmos
coscom (= 1)
3 = Printer protocol (serial printer or converter required)
4 = Free
5 = Pacer treadmill emulation (max. 12.7 km/h)
6 = MAX 1 (GE Marquette) EKG; if available use: h/p/cosmos
coscom (= 1)
7 = Trackmaster treadmill emulation in km/h
8 = Trackmaster treadmill emulation in miles per hour
9 = custo card ECG (MS DOS-Version); if available use: h/p/cosmos
coscom (= 1)
10 = Loop Back Test (special test plug required, available at
h/p/cosmos)
11 = SunTech Tango blood pressure monitor
12 = Remote Control Hardware Terminal MCU 4
(special hardware required)
13 = Burdick T 500 treadmill emulation
14 = Burdick T 600 treadmill emulation
17 = Chipcard reader Ergofit (special hardware required) – only
available at MCU4 (all firmware versions)
or MCU5 firmware < v1.03.x
18 = Chipcard reader Proxomed (special hardware required) – only
available at MCU4 (all firmware versions) or MCU5 firmware <
v1.03.x
20 = h/p/cosmos coscom v3 with Baudrate 115200 bps
OP21
RS232 interface protocol: COM 2
See descriptions above
OP23
RS232 interface protocol COM 4
§
indicates flashing: number of the RS232 interface protocol
$
and
%
reports:
OFF = RS232 not active / no protocol / interface deactivated
18 = Chip-card reader PROXOMED (special hardware required)
20 = h/p/cosmos coscom v3 / Baudrate 115200 bps
22 = Polar W:I:N:D: - System (weitere Einstellungen in Benutzer
Option OP19 und Administrator Option OP16)