![background image](http://html2.mh-extra.com/html/h-p-cosmos/mercury/mercury_user-manual_3085347157.webp)
technical data
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 157 of 213
h/
p/
cosmos st
ellar
h/
p/
cosmos st
ellar lt
h/
p/
cosmos quas
ar
h/
p/
cosmos quas
ar lt
h/
p/
cosmos st
ellar m
ed
h/
p/
cosmos st
ellar m
ed lt
h/
p/
cosmos quas
ar med
h/
p/
cosmos quas
ar med lt
h/
p/
cosmos pulsar
h/
p/
cosmos pulsar lt
h/
p/
cosmos pulsar 3p
h/
p/
cosmos pulsar 3p lt
Dimension of running surface
length x width
1700 x 650 mm
1900 x 650 mm
Dimension frame
length x width x height
2300 x 1050 x 1450 mm
(UserTe 150 mm width)
2500 x1050 x 1450 mm
(UserTe 150 mm width)
Track access height from floor
230 mm
Weight of device
360 kg
370 kg
370 kg
380 kg
400 kg
440 kg
Weight load limits to floors
(traffic load to DIN 1055 part 3)
6.4 kN/m²
6.1 kN/m²
In case of any further questions related to load limits onto floors, please ask our service department.
Max. permissible user weight
200 kg (deviation on request specially made)
Environmental conditions:
Transport and storage
temperature: –30...+50 °C
humidity: 0...95% - without condensation
barometric pressure: 700...1060hPa
Environmental conditions:
Operation
temperature: +10...+40 °C (deviation on request)
humidity: 30...70% - without condensation (up to 95% on request)
barometric pressure: 700...1060hPa
Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
Pollution degree
Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.
Central lubrication system
manual
Store program space
6 training profiles / 10 test profiles / 8 user definable profiles
RS232 Digital interface / serial port
(USB-converter optional)
COM1: standard;
COM2: optional;
COM3: optional (standard for service);
COM4: optional
COM1 and COM2: standard
COM3: optional (standard for
service);
COM4: optional
Heart rate measurement and heart
rate dependent endurance control
POLAR, 1 channel receiver, ECG precisely
POLAR W.I.N.D. with WearLink
transmission range approx. 10 m
PC software
h/p/cosmos para control
®
included
options: h/p/cosmos para graphics
®
and h/p/cosmos para analysis
®
h/p/cosmos para control
®
and
h/p/cosmos para graphics
®
included
Reverse belt rotation / downhill
1
not applicable
option
not applicable
option
standard
Safety Arch with harness***
option
Handrails options:
short,
long, crossbar (standard)
Arm support / additional keyboard
option
* Overload and weak power supply (undervoltage) could lead to higher discrepancies concerning the speed-display or dripping of the fuse. ** The time of
operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be
used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25%.
*** For maximum stress tests and with oversize running machines (track sizes) a safety facility (e.g. a safety harness with chest belt) is necessary and highly
recommended. When using a front-bar handrail (crossbar), the speed is limited to 5 km/h for safety reasons. Data of weight and measurements can deviate.
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.