safety, warnings, precautions, prohibitions
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 24 of 213
[4.E7]
Protection against the danger of electrical shocks
Casing and cover: Parts of non-medical
electric devices within the surrounding of
the subject, which, after having taken off
the coverings etc. without the use of tools
for a routine maintenance etc., are in
touchable reach, have to operate with a
voltage that does not exceed 25 volt
alternate voltage and 60 volt direct
voltage, which is produced by a separate
source as described in IEC 60601-1.
According to this example the leakage
current would flow from the electric device
to the earthened subject via the trainer/doctor.
Electrical equipment [3] and a subject [1] must never be touched by the doctor or trainer [2]
at the same time.
[4.E8]
Environmental requirements
Running machines are not to be used in medically utilized rooms with a danger of explosions or in easily
inflammable atmospheres. The devices must not be installed near to e.g. an x-ray device, motors or transformer
with high connection power, as the electric and magnetic interference can falsify measurements or even make
them impossible. High voltage lines must be avoided nearby the device. h/p/cosmos electrical devices with mains
connections must neither be used in wet and humid areas (e.g. swimming pool, sauna, etc.) nor in environmental
chambers.
If not stated otherwise in the delivery information h/p/cosmos devices are designed for operation in normal climatic
surroundings (IEC 60601-1):
N
Temperature: +
10° ... + 40° C
N
Relative humidity:
30 ... 70 % (non condensing!)
N
Air pressure:
700 ... 1060 mbar
N
Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
The running machine has to be protected from high humidity. Venting slots are not to be covered; otherwise it
would hinder the air circulation. Store the devices at a temperature of – 20° ... + 50° C.
All devices can be stored without power-connection and without operation for a period of 6 – 9 months. After this
period batteries in the device may be discharged and maintenance and/or a new setup/programming of the device
could be required by authorized h/p/cosmos technician.