optional settings
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 106 of 213
[8.B]
Administrator options
The administrator options are only accessible by h/p/cosmos staff and authorised service engineers. The access is only
possible with a special code which is not depicted here.
Changing an administrator option might cause serious problems if you are not familiar with all the details. Therefore the
h/p/cosmos service has to be contacted before doing any changes in the administrator options.
For optional settings on devices without UserTerminal an external UserTerminal or a connected PC with the software
h/p/cosmos para control is required.
[8.B1]
List of the administrator options
[8.B1]
Option
Description
Comment / Display
OP01
Set back of error messages E 20, E 21,
E 30, E 31, E 32, E 41, E 50 and E 51
The required maintenance work has to be performed before deleting
error messages. Confirmation in the display
§
with "donE"
This option only sets back the error message. If the error still exists,
the error message will appear again
OP02
Selection of the OEM-Code
By selecting the OEM code the functions and settings of the
UserTerminal will be adapted to the requirements of the listed OEM
partner.
Following OEM codes can be selected:
0 = h/p/cosmos (Standard)
1 = JAEGER / VIASYS / CARDINAL HEALTH
2 = Ergo-Fit (1997-98)
3 = Proxomed / Kardiomed
4 = KISTLER Gaitway
6 = COSMED
9 = SCHILLER
Not all companies listed above are or were OEM partners of
h/p/cosmos.
OP03
Selection of device type
Selection of the correct treadmill device type according to the „list of
device types” in the appendix. On the ladder-ergometer only the device
type 0.1 and 1.1 can be confirmed..
Note: Selection and confirmation of a device type sets back all user-
and administrator options to default.