Calibration
Resolving problems
It is unlikely that you will encounter problems during calibration. However, if you
are unable to meet a calibration objective at any point in the process you must take
corrective action and start the process again from the beginning.
If you encounter a problem-
Check your equipment, test cavities and connections, and start over.
Still have a problem-
Clean the probe or Contra phone, check the probe tubes and start over.
Still have a problem-
Change the probe assembly and start over.
Still have a problem-
Contact your local GSI representative or GSI technical support.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 27
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...