Calibration
Step 8
Adjust the overall calibration output level if necessary by pressing the Cal HL up/
down softkeys. The level can be adjusted up or down in 5.0dB increments.
Step 9 (Custom Mode only)
When a custom transducer is being calibrated as discussed earlier in Custom Trans-
ducers, the TARG SPL softkeys will be active.
Adjust the target SPL requirement for each frequency by pressing the TARG
SPL up/down softkeys. The level can be adjusted up or down in 0.5dB incre-
ments. The final values must be within 10dB of the GSI defaults.
Step 10
Follow the instructions on the screen until all stimulus calibrations are complete.
Step 11
Press the RETURN hardkey twice to return to the Main Calibration Menu.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 61
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...